<para>Some software components need to be installed before using this
...
...
@@ -50,15 +50,27 @@ Refer to the <i>LabVIEW Help</i> for more information about software requirement
<h4>3. Known Issues</h4>
<para>
To report issues or provide feedback about this instrument driver,
please send an email to <ahref="mailto:instrument.drivers@ni.com">instrument.drivers@ni.com</a>.
please send an email to <ahref="mailto:H.Brand@gsi.de">H.Brand@gsi.de</a>.
</para>
<para>
<br>Some erros occur without known reaqson. Maybe the TPG300's micro processor need more time than expected under unknown circumstances.
<UL>
<LI> Error -1073807339: VISA Timeout <br> TPG300 times out once in a while. Repeat command.</LI>
<LI> Error 85: Scan from String <br> Some commands seem to become misinterpreted once in a while. In such case the response can not be correctly interperted. An error number is maybe returned instead.</LI>
<LI> Error -1073807343: VISA Insufficient Location Information.</LI>
<LI> Error -1073807194: VISA Connection lost. <br> Previous two errors happen once in a while when using Wiesemann & Theis COM-Server.</LI>
</UL
</para>
<h4>4. Revision History</h4>
<para>The latest version of this and other LabVIEW and LabWindows/CVI instrument drivers can be downloaded at the <ahref="http://www.ni.com/idnet">Instrument Driver Network</a>.<br>
<para>The latest version of this instrument driver can be downloaded at the <ahref="https://git.gsi.de/EE-LV/Drivers/TPG300">git.gsi.de</a>.<br>
<br>
REV 2.0, 11/24/2017<br>
Modified by: H.Brand@gsi.de<br>
Improved communication. Refer to Git history.<br>
<br>
<fontcolor="Red">Provide detailed driver revision history including driver revision number, date of revision, and description of revision made to driver. Example:<br></font>