FROG Calibrator: Additionally save raw camera data user input in order to perform a post-analysis of the calibration
Instead of saving the spectral and temporal calibration results only, it would be very good to also save additional values / acquisitions in order to perform a post-analysis of the calibration:
Spectral calibration:
- Wavelengths used for the calibration
- Respective pixel cooresponding to the different wavelengths (raw value)
- Acquired image used for the calibation
Temporal calibration:
- Delay stage values (µm)
- Respective pixel corresponding to the maximum of the FROG trace in time-axis direction (raw value)
- Image for each acquired time-/ delay-step